EasyLog EL-USB-TC-LCD - Thermocouple Data Logger with LCD and USB Interface

EL-USB-TC-LCD,Thermocouple Data Logger with LCD and USB Interface,Thermocouple,Data,Logger,LCD,USB,Interface,EasyLogThis standalone data logger measures and stores up to 32,510 temperature readings from either a K, J or T type thermocouple.
A thermocouple is attached via the thermocouple socket at the base of the unit. The user can easily set up the logging rate and start-time, and download the stored data by plugging the data logger into a PC’s USB port and running the purpose designed software under Windows 2000, XP and Vista (32-bit). Data can then be graphed, printed and exported to other applications.
The high contrast LCD can show several different temperature variables. The user can cycle between these using the push button. The data logger is supplied complete with a long-life lithium battery, which can typically allow logging for up to 6 months.

EasyLog EL-USB-TC-LCD Thermocouple Data Logger Features

  • -200 to +1350ºC (-328 to +2462ºF) K-type measurement range
  • -200 to +1190ºC (-328 to +2174ºF) J-type measurement range
  • -200 to +390ºC (-328 to +734ºF) T-type measurement range
  • High contrast LCD, with four digit temperature display
  • USB Interface for set-up and data download
  • User-programmable alarm thresholds
  • Status indication via red and green LEDs
  • Immediate, delayed and push-to-start logging
  • Supplied with basic K-type thermocouple rated from 0 to 400°C (32 to 752°F)
  • Supplied complete with replaceable internal lithium battery and Windows control software

Windows Control Software

EL-USB-TC-LCD,Thermocouple Data Logger with LCD and USB Interface,Thermocouple,Data,Logger,LCD,USB,Interface,EasyLogEL-USB-TC-LCD,Thermocouple Data Logger with LCD and USB Interface,Thermocouple,Data,Logger,LCD,USB,Interface,EasyLogEL-USB-TC-LCD,Thermocouple Data Logger with LCD and USB Interface,Thermocouple,Data,Logger,LCD,USB,Interface,EasyLog

EasyLog USB control software is supplied free of charge with each data logger. Easy to install and use, the control software runs under Windows 2000, XP (Home and Professional Editions) & Vista (32-bit). The software is used to set-up the data logger as well as download, graph and export data to Excel. The software allows the following parameters to be configured:

  • Logger name
  • °C, °F
  • Logging rate (1s, 10s, 1m, 5m, 30m, 1hr, 6hr, 12hr)
  • High and low alarms
  • Immediate, delayed and push-to-start logging
  • Display off, on for 30 seconds after button press, or permanently on
  • Data rollover (Allows unlimited logging periods by overwriting the oldest data when the memory is full)

The latest version of the control software may be obtained free of charge contact hcs@hcs77.com

EL-USB-TC-LCD,Thermocouple Data Logger with LCD and USB Interface,Thermocouple,Data,Logger,LCD,USB,Interface,EasyLogYour Thermocouple

The probe supplied with your EL-USB-TC-LCD is a K-type thermocouple designed to measure temperatures from 0 to 400ºC (32 to 752ºF). The EL-USB-TC-LCD is designed to work with J, K and T-type thermocouples fi tted with a standard mini thermocouple connector; this makes your data logger compatible with a wide range of available thermocouples. Your application will determine which probe is most suitable based on temperature range, accuracy, form and price.

LED Flashing Modes

EL-USB-TC-LCD features two LEDs that indicate the logging, battery and alarm status:

  • The first LED fl ashes red (R) to indicate that the EL-USB-TC-LCD is in an alarm condition. It will flash when the logged temperature has exceeded a Low or High alarm level.
  • The second LED fl ashes green (G) to indicate that the EL-USB-TC-LCD is not in an alarm condition.

Hold is enabled by default, which forces the logger to continue flashing the red LED after an alarm, even when the temperature has returned to normal. This feature ensures that the user is notifi ed that an alarm level has been exceeded, without the need to download the data from the logger.

Hold can be turned off via the control software. The red LED will then only fl ash whilst the logger is in an alarm condition. When the temperature returns to normal, the green LED will flash.

EasyLog Main | EL-USB-TC-LCD | EL-USB-2-LCD | EL-2 | EL-USB-3 | EL-USB-4

For Quick Delivery From Stock Call
CANADA & USA TOLL FREE: 1 - 800 - 279-9912